Company record
Cadwell Industries Inc
0 FDA warning letters since 2023 and 1 recall in the last 12 months.
Get an email the day FDA posts anything new about Cadwell Industries Inc.
Recalls
| Reported | Class | Firm | Product and reason |
|---|---|---|---|
| Oct 21 | Class II | Cadwell Industries Inc Device |
Brand Name: IOMAX Product Name: IOMAX Cortical Module Model/Catalog Number: 190296-200 UDI-DI code: 00840067100653 Product Description: The IOMAX Cortical Module contains a 16-channel amplifier, a low current electri… Stimulator device contains a defective printed circuit board assembly (PCBA) that may cause the device to unintentionally deliver electrical stimulation to the patient, or allow an unintended electrical current to reach the patient. |